HVeVR2

SuperCDMS Public Documentation

HVeV Run 2


Abstract This Public Documentation accompanies the results on electron-recoiling dark matter (ERDM) from a 3 week long run of a 1g HVeV detector in the Northwestern ADR. During this run, one week of data (roughly 12 hours per day for 7 days) was acquired at 100V bias with a charge resolution of 0.03 electron-hole pairs, and 3 days at 60V with a resolution of 0.05 electron-hole pairs. Co-57 and laser calibrations were acquired, and a 0V control sample was also acquired over multiple days. The resulting limits for DM-electron scattering and DM absorption are given below, along with plots approved for public release. 
These results are availiable from arXiv:2005.14067


 

All results assume a relic DM density of ρDM = 0.3 GeV/cm3. All limits are at 90% confidence level. σe is the DM-electron scattering cross section. FDM is the DM form factor where FDM = 1 (FDM ∝ 1/q2) corresponds to a DM model with a heavy (ultra light) dark photon mediator. q is the momentum transfer. mDM is the DM mass. More detailed information on the assumed DM models and astrophysical parameters can be found in the US Cosmic Visions Community Report.  
 

Result


Topic


Last Updated


Comments/Download


HVeV Detector

Experimental Setup

August 16th, 2019

Photograph of the HVeV R2 detector
Picture of the HVeV R2 detector (left) with side RF veto (right) inside of the copper housing.

 

png, pdf

HVeV Detector Lateral View

Experimental Setup

January 30th, 2020

Photograph of the HVeV R2 detector
Picture of the HVeV detector seen from the side

 

png, pdf

Detector in Animal

Experimental Setup

January 30th, 2020

Photograph of the HVeV R2 detector installed in the ADR

Picture of the HVeV detector installed in the ADR at Northwestern University. (ArXiv:1903.06517). The ADR has two salt pills, GGG operating around 300 mK and FAA operating around 50 mK. The detector is thermally linked to the FAA salt, with temperature stabilized to 50 mK or 52 mK during the operation. A superconducting magnetic field shield, the Nb can, is mounted on the FAA stage around the detector box. The cold electronic readout device, SQUIDs, are mounted on a thermal stage operated at 1.3 K.

 

jpg, pdf

ADR Lab

Experimental Setup

January 30th, 2020

Photograph of the experimental set-up at the Northwestern University

Photograph of the experimental set-up at the Northwestern University, including the ADR, the computers controlling the cryostat, the data acquisition system.

 

jpg, pdf

TES mask drawing

Experimental Setup

February 21st, 2020

Drawing of the TES mask

Drawing of the sensor mask used for the HVeV R2 detector. Two channels with the same-area are visible and their contacts highlighted with darker squares.

 

png, pdf

Exposure Figure

Data Acquisition

February 21st, 2020

Exposure figure
Summary of total exposure as a function of wall time during the run by data type.

 

png,pdf

HVeVR2 Phonon Energy Resolution

Detector Response

February 21st, 2020

Phonon resolution on each day of acquisition

Phonon resolution as measured using laser calibration data for each day of Run 2. Variations are due to noise changes and change in TES bias point day to day. With a few exceptions, we consistently achieve a resolution of ~3.2 eV in the first electron-hole peak. The first point is inconsistent with the others because of an excessively noisy power supply that was replaced starting day 2.

 

png, pdf

HVeVR2 Resolution Change

Detector Response

February 21st, 2020

Charge resolution on each day of acquisition

Charge resolution as measured using laser calibration data for each day of Run 2. Variations are due to changes in noise, but largely are determined by the voltage bias used for each calibration as expected due to NTL gain. Resolution for points at the same bias are consistent throughout the run. The first point is inconsistent with the others because of an excessively noisy power supply that was replaced starting day 2.

 

png, pdf

HVeVR2 Relative Rescaling

Detector Response

February 21st, 2020

Relative gain.

Scatter plot comparing the inner and the outer channel for one day of background data: both channel have the same gain thanks to the same-area sensor design of the TES mask. The two anti-diagonal dashed lines indicate the first (lower left) and the second (upper right) electron-hole pair peaks.

 

png, pdf

Temperature Monitor

Detector Response

February 21st, 2020

Convertion between temperature monitor reading and fridge temperature.

The ADR temperature was monitored during all the runs and recoded each second. The temperature value was recorded in volt which can be then converted with the law reported in this figure.

 

png, pdf

HVeVR2 Temperature Summary

Detector Response

February 21st, 2020

Base temperature during data taking as a function of time.

Temperature was stabilized at 50 mK during the run for the first 9 days, after which it was raised to 52 mK in order to increase hold time. Periodic deviations in temperature can be seen, as well as rises in temperature at the end of each day as the fridge runs out of cooling power. Light/yellow-ish colors refer to a high count density and dark/purple colors refer to a low count density.

 

png, pdf

HVeVr2 Temperature Monitor

Detector Response

February 21st, 2020

Laser amplitude as a function of the temperature for twelve series acquired at different temperatures.

We acquired twelve laser series at different temperatures in order to correct for gain variations due to temperature instabilities and drifts. This plot shows the peak position fitted from the laser data as a function of the temperature.

 

png, pdf

HVeVR2 Variation

Detector Response

February 21st, 2020

Spread of the HV applied on the detector for different series.

The HV was set each day by hand with a knob and it was recorded each second by the DAQ. We found differences between the average values set each day and we corrected for them in the HV correction. The uncertainties are present in this plot but they are so small that are not visible.

 

png, pdf

HVeVR2 Linearity 100V

Detector Response

February 21st, 2020

Linearity correction and calibration.

The linearity correction and calibration of the HVeV R2 detector was done with the laser data acquired daily. This plot - showing the energy as a function of the optimum filter amplitude - is an example of the curve used for the calibration and linearity correction in one day.

 

png, pdf

HVeVR2 Linearity Test 100V

Data

February 21st, 2020

Combined laser spectra after the corrections and calibration for the 100 V data

Top panel: combined spectra of laser data (black histogram) at 100 V after the corrections (temperature, high voltage, QET absorption, linearity) and the calibration. The red curves represent Gaussian fits to each individual peaks. The expected peak positions are multiples of the photon energy (1.95 eV) plus the NTL gain energy for 1 electron-hole pair (100 eV). Bottom panel: residuals between the laser peak position after the calibration and the expected value.

 

png, pdf

HVeVR2 Linearity Test 60V

Data

February 21st, 2020

Combined laser spectra after the corrections and calibration for the 60 V data

Top panel: combined spectra of laser data (black histogram) at 60 V after the corrections (temperature, high voltage, QET absorption, linearity) and the calibration. The red curves represent Gaussian fits to each individual peaks. The expected peak positions are multiples of the photon energy (1.95 eV) plus the NTL gain energy for 1 electron-hole pair (60 eV). Bottom panel: residuals between the laser peak position after the calibration and the expected value.

 

png, pdf

Live time Temperature

Livetime cuts

February 21st, 2020

Example of one of the livetime cuts: temperature cut for the 50 mK data.

This histogram shows the temperature distribution before and after the temperature livetime cut for the 50 mK data.

 

png, pdf

Live time chart comparing 100V and 60V

Livetime cuts

February 21st, 2020

Pie chart of the livetime cuts for the 60 V and 100 V data.

This pie chart shows the percentage of data accepted and removed by the livetime cuts: temperature cut, mean baseline cut and burst cut.

 

png, pdf
png, pdf

HVeVR2 Laser data at 100v

Data quality cuts

February 21st, 2020

Frequency-domain Χ2 cut for the 100 V data

The frequency-domain Χ2 cut belongs to the data quality cuts and was studied with the laser data. This plot shows the frequency-domain Χ2 as a function of the energy. The orange dots represent the peak of the Χ2 distribution for each electron-hole pair peak. The red dots represent 3σ position of the corresponding peak. The red line corresponds to the position of the cut.

 

png, pdf

HVeVR2 Time Domain 100V

Data quality cuts

February 21st, 2020

Time-domain Χ2 cut for the 100 V data

The time-domain Χ2 cut belongs to the data quality cuts and was studied with the laser data. This plot shows the time-domain Χ2as a function of the energy. The orange dots represent the peak of the Χ2 distribution for each electron-hole pair peak. The red dots represent 3σ position of the corresponding peak. The red line corresponds to the position of the cut.

 

png, pdf

HVeVR2 Cumulative Data

Data quality cuts

February 21st, 2020

Cumulative data quality cuts for laser and dark-matter-search data at 100 V

This histogram shows the laser and dark-matter-search spectrum after the application of the data-quality cuts.

 

png, pdf
png, pdf

HVeVR2 Triggered Data

Trigger

February 21st, 2020

Trigger efficiency for the 100 V data

Top panel: comparison of the combined laser spectrum at 50 mK triggered with the laser TTL and triggered with the trigger used for the dark-matter-search data. Bottom panel: ratio of the histograms reported on the top panel, the trigger theshold at 50 mK is 29 eV. For this analysis, the lower limit of the region of interest was set at 50 eV to simplify the treatment of trigger effects. Note that the same procedure was applied to the 52 mK data and we found a threshold of 37 eV.

 

png, pdf

Efficieny Data 100V

Efficiency

February 21st, 2020

Cut Efficiency for the 100 V data

Efficiency as a function of the energy for the 100 V data. The efficiency is fit with the red curve and the fit uncertainty is highlighted with the band. The fit curve and its uncertainty are used to set the limit.

 

png, pdf

Efficiency 60V

Efficiency

February 21st, 2020

Cut Efficiency for the 60 V data

Efficiency as a function of the energy for the 60 V data. The efficiency is fit with the red curve and the fit uncertainty is highlighted with the band. The fit curve and its uncertainty are used to set the limit.

 

png, pdf

Laser Spectrum 100V

Data

March 27st, 2020

Cumulative laser calibration spectrum with cuts applied with 100 V data

Laser calibration data for all 100 V data. The data are plotted in three cases: (1) after the corrections and the calibration; (2) after the livetime cuts; (3) the data quality cuts.

 

png, pdf

More HVeV Run 2

 

References for Fig. 30 and 31:


References for Fig. 32 and 33:


References for Fig. 34:

  1. G. Hildebrandt et al. Normale und anomale Absorption von Röntgen-Strahlen in Germanium und Silicium/Normal and anomalous absorption of X-rays in Germanium and Silicon. Z. Naturforsch. A, 28:588, 1973.
  2. F. C. Brown and O. P. Rustgi. Extreme Ultraviolet Transmission of Crystalline and Amorphous Silicon. Phys. Rev. Lett., 28:497, 1972.
  3. C. Gahwiller and F. C. Brown. Photoabsorption near the LII, III Edge of Silicon and Aluminum. Phys. Rev. B, 2:1918, 1970.
  4. B. L. Henke et al. X-Ray Interactions: Photoabsorption, Scattering, Transmission, and Reflection at E =50-30,000 eV, Z = 1-92. At. Data Nucl. Data Tables, 54:181, 1993.
  5. W. Da-Chun et al. Measurement of the mass attenuation coefficients for SiH4 and Si. Nucl. Instrum. Meth. B., 95:161, 1995.
  6. M. A. Green and M. J. Keevers. Optical properties of intrinsic silicon at 300 K. Prog. Photovoltaics, 3:189, 1995.
  7. M. A. Green. Self-consistent optical parameters of intrinsic silicon at 300 K including temperature coefficients. Sol. Energy Mater. Sol. Cells, 92:1305, 2008.
  8. W. C. Dash and R. Newman. Intrinsic Optical Absorption in Single-Crystal Germanium and Silicon at 77 oK and 300 oK. Phys. Rev., 99:1151, 1955.
  9. D. F. Edwards.- Silicon (Si)*. In Edward D. Palik editor, Handbook of Optical Constants of Solids, pages 547 � 569. Academic Press, Burlington, 1997.
  10. G. G. Macfarlane et al. Fine Structure in the Absorption-Edge Spectrum of Si. Phys. Rev., 111:1245, 1958.
  11. S. E. Holland et al. Fully-depleted, back-illuminated charge-coupled devices fabricated on high-resistivity silicon. IEEE Trans. Electron Devices, 50:225, 2003.
  12. E. M. Gullikson et al. Absolute photoabsorption measurements of Mg, Al, and Si in the soft-x-ray region below the L2,3 edges. Phys. Rev. B, 49:16283, 1994.
  13. M. J. Berger et al. XCOM: Photon cross section database (version 1.5), 2010.
  14. F. C. Brown et al. L2,3 threshold spectra of doped siliconand silicon compounds. Phys. Rev. B, 15:4781, 1977.
  15. W. R. Hunter. Observation of Absorption Edges inthe Extreme Ultraviolet by Transmittance Measurements Through Thin Unbacked Metal Films.
  16. In F. Abelés, editor, Optical Properties and Electron Structure of Metals and Alloys, pages 136�146. North-Holland Publishing Company - Amsterdam, 1966.